Instructions

Assembly Instructions
  1. Hold respirator with silicone gasket open to your face.
  2. Place Diverg disposable filters on the trim edge of the outer compression ring.
    NOTE: This respirator should be used only with DIVERG disposable filters.
  3. Push one side of the filter trim to secure into the slot. Then push the other side. With both fingers, secure the rest of the seal.
  4. Pull strap and check for a firm fit.
Fitting Instructions
Must be followed each time respirator is worn.
DONNING:
  1. Place the assembled respirator on nose and mouth.
  2. Pull straps over the head.
  3. Position respirator low on the bridge of the nose to ensure optimal visibility and best fit.
  4. Adjust straps for optimal tightness.
DOFFING:
  1. Loosen the adjustable straps of the respirator.
  2. Take the straps off from above the head.
  3. Remove respirator.
Fitting Testing
The effectiveness of the respirator will be reduced if it is not properly fitted. As per OSHA and Canadian CSA requirements, either qualitative or quantitative fit testing should be done prior to use.
Replacement Filters
Replacement filters should be stored away from any contaminants / contaminated areas when not in use.
Filter Disposal
Filters might contain hazardous particles after use. Should be disposed of following the local government regulations for hazardous material.
Cleaning, Inspection and Storage
  1. The following cleaning instructions are made using OSHA and CSA guidelines
  2. Cleaning is recommended after each use. In Healthcare settings, disinfection frequency will be based on the facility infection control policy but should also reference guidance from OSHA that states "...respirators must be cleaned as often as necessary to prevent them from becoming unsanitary. In addition, respirators worn by more than one user must be cleaned and disinfected before being worn by a different user."
  3. Nitrile or vinyl gloves should be worn during cleaning as well as other personal protective equipment (PPE) as indicated.
  4. Disassemble the respirator by removing the filter, shell and straps. Discard the filter.
  5. Inspect all the components per the inspection instruction to identify any damage or overuse. As per OSHA regulations, inspection shall include the following:
    • Conditions of component parts (i.e., shell, seal and disposable filter)
    • Tightness of connectors
    • End-of-service-life indicator
    • Shelf-life
    • Proper functioning of regulators, alarms and warning systems
  6. Manual cleaning: Wash with warm water not exceeding 43C (110F), using soft brush to thoroughly clean all surfaces. Use mild detergents as cleansers. Solvents/strong detergents are not recommended as they may cause deterioration of the respirator components. Rinse thoroughly with warm water until residual cleansers are washed off.
  7. Disinfection: when the cleaner used does not contain a disinfecting agent, respirator components should be immersed for two minutes in one of the following (Per OSHA's guidelines):
    • Hypochlorite solution(50ppm of chlorine) made by adding approximately one milliliter of laundry bleach to one liter of water at 43C(110F)
    • Aqueous solution of iodine(50ppm iodine) made by adding approximately 0.8milliliters of tincture of iodine(6-8grams ammonium and/or potassium iodide/100cc of 45% alcohol)to one liter of water at 43C(110F)
  8. Rinse components thoroughly in clean, warm and preferably running water until all residual disinfectants are washed off. [NOTE:Residual disinfectants may cause deterioration of the respirator components. Through rinsing is important.]
  9. Hand dry components with a clean, lint-free cloth or air dry.
  10. Reassemble respirator following the assembly instructions.
  11. Test respirator before use to ensure all components work properly.
  12. This product can be used for the maximum of 200 cleaning cycles.
  13. Clean respirator should be stored away from any contaminants/contaminated area when not in use.